CVE-2014-5406
NONEThe Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote
Опубликовано: 7/6/2015Обновлено: 11/3/2025
Описание
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
ИИ-АнализНа базе ИИ
Затронутые продукты
hospiralifecare_pcainfusion_firmware
hospiralifecare_pca3
-
hospiralifecare_pca5
-
Ссылки
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htmThird Party AdvisoryUS Government Resource
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htmThird Party AdvisoryUS Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01Third Party AdvisoryUS Government Resource
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/